T4/T6 VNA

T4/T6 VNA

$4,950.00$6,450.00

USB Vector Network Analyzer

T series USB Vector Network Analyzer offers wide dynamic range,low noise level, high resolution scanning with laboratory and research grade performance. T4 covers frequency range of 1MHz to 4GHz and T6 covers frequency range of 1MHz to 6GHz, with 2-port and 2-pass that competitive with most of the bench-top VNAs on the market. The T series USB VNA provides measurement convenience by offering end user excellent performance and attractive pricing. T series USB VNA is suitable for laboratory, manufacturing and many other safety testing environment.

Clear

SKU: N/A.
Category: .
  • Frequency Range:1MHz to 4GHz (T4) 500kHz to 6.5GHz (T6)
  • Dynamic Range:>120 dB (IFBW=10 Hz), 123dB typical
  • Low Noise Level:<-120 dB (IFBW=10 Hz)
  • Low Trace Noise:10 mdB rms (IFBW=3 kHz)
  • High Measurement Speed:100 μs/point (IFBW=30 kHz)
  • High Effective Directivity:>42 dB
  • Remote Control:LabView
  • Capable of replacing bench-top VNA
  • Minimum budget requirement
  • Suitable for laboratory, manufacturing and research and development purposes
  • Compact design, implementation simplicity and various system upgrade

Testing Range                                                                

Description

Impedance

50Ω

Test Port Connector

N-type, female

Number of Test Port

2

Frequency Range

T4: 100MHz to 4GHz;  (Low range option1 to 100MHz)

T6: 500KHz to 6.5GHz;  

Frequency Accuracy

5ppm

Frequency Resolution

10Hz

Number of Measurement Points

2 to 10001

Measurement Bandwidths

10Hz to 100kHz

Dynamic range (IFBW 10Hz)

120dB, typ.123 dB

Measurement parameters

S11, S21, S12, S22

Testing Accuracy

Transmission Measurement Accuracy (magnitude/phase)

 

+5 dB to +10 dB

0.2dB/2º

-50 dB to +5 dB

0.1dB/1º

-70dB to -50dB

0.5 dB/3º

-90dB to -70dB

2.5 dB/11º

Reflection Measurement Accuracy(magnitude/phase)

 

-15dB to 0dB

0.4dB/3º

-25dB to -15dB

1.0dB/6º

-35dB to -25dB

3.0dB/20º

Trace stability

 

Trace noise magnitude (IF bandwidth 3kHz)

10 mdB rms

Temperature Dependence (per one degree of temperature variation)

0.02dB

Effective System Data①

Effective Directivity

42 dB

Effective Source Match

40 dB

Effective Load Match

42 dB

①Applied over them temperature range of 23°C ± 5°C after 40 minutes of warming-up, with less than 1°deviation from the full two-port calibration temperature, at output power of -5dBm and IF bandwidth 10Hz.

 

Test port output                                                             

Description

Match(W/O system error correction)

18 dB

Power Range

-30 dBm to +5 dBm

Power Accuracy

±1.5 dB

Power Resolution

0.05dB

Test port input

Match(W/O system error correction)

18 dB

Max Input Level

+23 dBm

Max Input Voltage

+35 V

Noise level(IF Bandwidth 10Hz)

-115 dBm

General Data

External Reference Input

SMA female; 10 MHz; 2 dBm ± 3 dB

External Reference Output

SMA female; 10 MHz; 3 dBm ± 2 dB

Operating Temperature Range

+5°C to +40°C

Storage Temperature Range

-45°C to +55°C

Operating Humidity

90% (25°C)

Operating Atmospheric Pressure

84 to 106.7 kPa

Calibration Interval

3 years

Power Supply

110/220 ± 22 V (AC), 50 Hz

 

Comparision between VNA modules

 

Transcom

T series

Tektronix

TTR500

Copper Mountain

S5048

Frequency Range

1MHz to 4GHz (T4)

500kHz to 6.5GHz (T6)

100kHz to 3GHz

100kHz to 6GHz

20kHz to 4.8GHz

Dynamic Range(IFBW 10 Hz)

120dB

122dB

120dB

Trace Noise (dBrms)

0.01

0.008

0.002

Output Power

-30 to +5 dBm

-50 to +7 dBm

-50 to +5 dBm

Measurements

S11,S12,S21,S22

S11,S12,S21,S22

S11,S12,S21,S22

Frequency Resolution

10Hz

1Hz

10Hz

Frequency Accuracy

±5ppm

±7ppm

±5ppm

Model                                                                            

Description

Main Unit

T4/T6

2 port USB vector network analyzer 

Option Unit

LF-1

1M low frequency range option

Recommended Accessories